BS-6955-17-1994 ISO-5347-17-1993.pdf

上传人:来看看 文档编号:3775676 上传时间:2019-09-23 格式:PDF 页数:8 大小:302.88KB
返回 下载 相关 举报
BS-6955-17-1994 ISO-5347-17-1993.pdf_第1页
第1页 / 共8页
BS-6955-17-1994 ISO-5347-17-1993.pdf_第2页
第2页 / 共8页
BS-6955-17-1994 ISO-5347-17-1993.pdf_第3页
第3页 / 共8页
BS-6955-17-1994 ISO-5347-17-1993.pdf_第4页
第4页 / 共8页
BS-6955-17-1994 ISO-5347-17-1993.pdf_第5页
第5页 / 共8页
亲,该文档总共8页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

《BS-6955-17-1994 ISO-5347-17-1993.pdf》由会员分享,可在线阅读,更多相关《BS-6955-17-1994 ISO-5347-17-1993.pdf(8页珍藏版)》请在三一文库上搜索。

1、BRITISH STANDARD BS 6955-17: 1994 ISO 5347-17: 1993 Calibration of vibration and shock pick-ups Part 17: Method of test for fixed temperature sensitivity UDC 534.1:681.327.7:53.089.6 Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 05:17:41 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6955-17

2、:1994 This British Standard, having been prepared under the direction of the General Mechanical Engineering Standards Policy Committee, was published under the authority of the Standards Board and comes into effect on 15 March 1994 BSI 11-1999 The following BSI references relate to the work on this

3、standard: Committee reference GME/21 Draft for comment 87/72304 DC ISBN 0 580 23089 9 Committees responsible for this British Standard The preparation of this British Standard was entrusted by the General Mechanical Engineering Standards Policy Committee (GME/-) to Technical Committee GME/21, upon w

4、hich the following bodies were represented: Electricity Association Federation of Civil Engineering Contractors Imperial College of Science and Technology Institute of Sound and Vibration Research Institution of Mechanical Engineers Lloyds Register of Shipping Ministry of Defence Open University Pow

5、er Generation Contractors Association PGCA (BEAMA Ltd.) Railway Industry Association of Great Britain Sira Test and Certification Ltd. Society of British Aerospace Companies Limited The following bodies were also represented in the drafting of the standard, through subcommittees and panels: British

6、Coal Corporation Health and Safety Executive Society of Environmental Engineers University of Cranfield Amendments issued since publication Amd. No.DateComments Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 05:17:41 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6955-17:1994 BSI 11-1999i Con

7、tents Page Committees responsibleInside front cover National forewordii 1Scope1 2Normative reference1 3Apparatus1 4Preferred amplitudes, frequencies and temperatures1 5Method1 List of referencesInside back cover Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 05:17:41 GMT+00:00 2006, Uncontro

8、lled Copy, (c) BSI BS 6955-17:1994 ii BSI 11-1999 National foreword This Part of BS 6955 has been prepared under the direction of the General Mechanical Engineering Standards Policy Committee. It is identical with ISO 5347-17:1993 Methods for the calibration of vibration and shock pick-ups Part 17:

9、Testing of fixed temperature sensitivity, published by the International Organization for Standardization (ISO). ISO 5347-17 was prepared by Technical Committee ISO/TC 108, Mechanical vibration and shock, in which the UK played an active part. BS 6955 consists of the following Parts, which are ident

10、ical with the corresponding Parts of ISO 5347: Part 0: Guide to basic principles; Part 1: Methods for primary vibration calibration by laser interferometry; Part 2: Method for primary shock calibration by light cutting; Part 3: Method for secondary vibration calibration; Part 4: Method for secondary

11、 shock calibration; Part 5: Method for calibration by Earths gravitation; Part 6: Method for primary vibration calibration at low frequencies; Part 7: Methods for primary calibration by centrifuge; Part 8: Method for primary calibration by dual centrifuge; Part 9: Method for secondary vibration cali

12、bration by comparison of phase angles; Part 10: Method for primary calibration by high impact shocks; Part 11: Method of test for transverse vibration sensitivity; Part 12: Method of test for transverse shock sensitivity; Part 13: Method of test for base strain sensitivity; Part 14: Method of test f

13、or resonance frequency of undamped accelerometers on a steel block; Part 15: Method of test for acoustic sensitivity; Part 16: Method of test for mounting torque sensitivity; Part 17: Method of test for fixed temperature sensitivity; Part 18: Method of test for transient temperature sensitivity; Par

14、t 19: Method of test for magnetic field sensitivity. Part 20 of ISO 5347 is in preparation. It is envisaged that when it is published it will be implemented as Part 20 of BS 6955. A British Standard does not purport to include all the necessary provisions of a contract. Users of British Standards ar

15、e responsible for their correct application. Compliance with a British Standard does not of itself confer immunity from legal obligations. Cross-reference International StandardCorresponding British Standard ISO 5347-3:1993BS 6955 Calibration of vibration and shock pick-ups Part 3:1994 Methods for s

16、econdary vibration calibration (Identical) Summary of pages This document comprises a front cover, an inside front cover, pages i and ii, pages 1 and 2, an inside back cover and a back cover. This standard has been updated (see copyright date) and may have had amendments incorporated. This will be i

17、ndicated in the amendment table on the inside front cover. Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 05:17:41 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6955-17:1994 BSI 11-19991 1 Scope ISO 5347 comprises a series of documents dealing with methods for the calibration of vibration an

18、d shock pick-ups. This part of ISO 5347 lays down detailed specifications for the instrumentation and procedure to be used for fixed temperature sensitivity testing. It applies to rectilinear pick-ups. This part of ISO 5347 is applicable for the following parameters: frequency range: 20 Hz to 1 250

19、Hz; dynamic range: 0,1 4m to 10 mm (frequency-dependent); 1 mm/s to 1 m/s (frequency-dependent); 10 m/s2 to 1 000 m/s2 (frequency-dependent); temperature range: 45 C to + 800 C. The uncertainty applicable is 10 % of reading. 2 Normative reference The following standard contains provisions which, thr

20、ough reference in this text, constitute provisions of this part of ISO 5347. At the time of publication, the edition indicated was valid. All standards are subject to revision, and parties to agreements based on this part of ISO 5347 are encouraged to investigate the possibility of applying the most

21、 recent edition of the standard indicated below. Members of IEC and ISO maintain registers of currently valid International Standards. ISO 5347-3:1993, Methods for the calibration of vibration and shock pick-ups Part 3: Secondary vibration calibration. 3 Apparatus 3.1 Equipment capable of maintainin

22、g room temperature at 23 C 3 C. 3.2 Vibration equipment, complying with the equipment specified for secondary vibration calibration (see ISO 5347-3). The vibrator shall have a fixture with thermal barrier so the test pick-up can be placed inside a temperature chamber and the reference pick-up outsid

23、e the chamber at laboratory conditions. The temperature influence on the reference pick-up and the fixture transfer function error shall be less than 0,5 % of reading. 3.3 Temperature chamber, designed so that the air temperature in the working space is evenly constant within 3 C for temperatures be

24、tween 65 C and + 100 C and within 5 C above 100 C. The temperature of the chamber walls shall not differ from the test temperature by more than 10 C at any point. Forced air circulation is recommended. The temperature sensors shall be protected against heat radiation and shall measure the air temper

25、ature before the air reaches the pick-up. 3.4 Sensor, for measuring the pick-up temperature, protected against heat radiation. The sensor shall be mounted on the pick-up or, if that is not feasible, on the heat-conducting plate. 4 Preferred amplitudes, frequencies and temperatures Six levels equally

26、 covering the pick-up range shall be chosen from the following series. a) Amplitude, in metres per second squared: 1; 2; 5; 10; and their multiples of ten; b) Frequency, in hertz: 20; 40; 80; 160; 315; 630; 1 250; NOTE 1In cases where a connecting rod is used between the vibrator and temperature cha

27、mber, care should be taken to avoid test frequencies at which high transverse motion of the rod occurs. c) Temperature, in degrees Celsius: 65; 50; 40; 25; 10; 0; + 5; + 40; + 70; + 100; + 155; + 200; + 250; + 400; + 800. Preferred values: 25 and + 70. 5 Method 5.1 Test procedure Calibrate the test

28、pick-up in the temperature chamber at laboratory temperature in accordance with a calibration method, usually the secondary vibration calibration (see ISO 5347-3). Check calibration frequencies and amplitudes so that limits given for distortion, transverse, bending and rocking acceleration, hum, noi

29、se and relative motion are complied with for the two pick-ups. Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 05:17:41 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6955-17:1994 2 BSI 11-1999 Then adjust the temperature within the chamber to the specified value and maintain this temperature

30、until the specimen has reached temperature stability, i.e. when the ratio of two consecutive time intervals necessary in order to change the temperature of the pick-up by 3 C exceeds a factor of 1,7. Record the temperature of the reference accelerometer casing. Then measure the calibration factor at

31、 specified frequencies and amplitudes. 5.2 Expression of results Estimate the temperature sensitivity as a percentage deviation of the calibration factor relative to the calibration factor at laboratory temperature and at reference frequency and amplitude. Calculate the calibration factor and the un

32、certainty in accordance with the calibration method chosen (see the relevant part of ISO 5347). Errors caused by the temperature influence on the reference pick-up and the fixture transfer function error which have not been compensated for shall be added to the errors in the relevant uncertainty cal

33、culations. Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 05:17:41 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6955-17:1994 BSI 11-1999 List of references See national foreword. Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 05:17:41 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 69

34、55-17: 1994 ISO 5347-17: 1993 BSI 389 Chiswick High Road London W4 4AL BSI British Standards Institution BSI is the independent national body responsible for preparing British Standards. It presents the UK view on standards in Europe and at the international level. It is incorporated by Royal Charte

35、r. Revisions British Standards are updated by amendment or revision. Users of British Standards should make sure that they possess the latest amendments or editions. It is the constant aim of BSI to improve the quality of our products and services. We would be grateful if anyone finding an inaccurac

36、y or ambiguity while using this British Standard would inform the Secretary of the technical committee responsible, the identity of which can be found on the inside front cover. Tel: 020 8996 9000. Fax: 020 8996 7400. BSI offers members an individual updating service called PLUS which ensures that s

37、ubscribers automatically receive the latest editions of standards. Buying standards Orders for all BSI, international and foreign standards publications should be addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001. In response to orders for international standards, it is BSI poli

38、cy to supply the BSI implementation of those that have been published as British Standards, unless otherwise requested. Information on standards BSI provides a wide range of information on national, European and international standards through its Library and its Technical Help to Exporters Service.

39、 Various BSI electronic information services are also available which give details on all its products and services. Contact the Information Centre. Tel: 020 8996 7111. Fax: 020 8996 7048. Subscribing members of BSI are kept up to date with standards developments and receive substantial discounts on

40、 the purchase price of standards. For details of these and other benefits contact Membership Administration. Tel: 020 8996 7002. Fax: 020 8996 7001. Copyright Copyright subsists in all BSI publications. BSI also holds the copyright, in the UK, of the publications of the international standardization

41、 bodies. Except as permitted under the Copyright, Designs and Patents Act 1988 no extract may be reproduced, stored in a retrieval system or transmitted in any form or by any means electronic, photocopying, recording or otherwise without prior written permission from BSI. This does not preclude the

42、free use, in the course of implementing the standard, of necessary details such as symbols, and size, type or grade designations. If these details are to be used for any other purpose than implementation then the prior written permission of BSI must be obtained. If permission is granted, the terms may include royalty payments or a licensing agreement. Details and advice can be obtained from the Copyright Manager. Tel: 020 8996 7070. Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 05:17:41 GMT+00:00 2006, Uncontrolled Copy, (c) BSI

展开阅读全文