《ISO-2884-1-1999.pdf》由会员分享,可在线阅读,更多相关《ISO-2884-1-1999.pdf(10页珍藏版)》请在三一文库上搜索。
1、A Reference number ISO 2884-1:1999(E) INTERNATIONAL STANDARD ISO 2884-1 First edition 1999-10-01 Paints and varnishes Determination of viscosity using rotary viscometers Part 1: Cone-and-plate viscometer operated at a high rate of shear Peintures et vernis Dtermination de la viscosit au moyen de vis
2、cosimtres rotatifs Partie 1: Viscosimtre cne et plateau gradient de vitesse lev ISO 2884-1:1999(E) ISO 1999 All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and mic
3、rofilm, without permission in writing from the publisher. International Organization for Standardization Case postale 56 CH-1211 Genve 20 Switzerland Internetisoiso.ch Printed in Switzerland ii Foreword ISO (the International Organization for Standardization) is a worldwide federation of national st
4、andards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International o
5、rganizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization. International Standards are drafted in accordance with the rules giv
6、en in the ISO/IEC Directives, Part 3. Draft International Standards adopted by the technical committees are circulated to the member bodies for voting. Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote. International Standard ISO 2884-1 w
7、as prepared by Technical Committee ISO/TC 35, Paints and varnishes, Subcommittee SC 9, General test methods for paints and varnishes. Together with the other parts (see below), this part of ISO 2884 cancels and replaces ISO 2884:1974, which has been technically revised. The main changes are: a) only
8、 a cone-and-plate viscometer is now specified; b) the shear rate is within a narrower range. ISO 2884 consists of the following parts, under the general title Paints and varnishes Determination of viscosity using rotary viscometers: Part 1: Cone-and-plate viscometer operated at a high rate of shear
9、Part 2: Disc or ball viscometer operated at a single specified speed Part 3: Disc and spindle viscometer operated at different speeds Part 4: Paddle viscometer operated at a single specified speed Annex A of this part of ISO 2884 is for information only. -,-,- INTERNATIONAL STANDARD ISOISO 2884-1:19
10、99(E) 1 Paints and varnishes Determination of viscosity using rotary viscometers Part 1: Cone-and-plate viscometer operated at a high rate of shear 1 Scope This part of ISO 2884 is one of a series dealing with the sampling and testing of paints, varnishes and related products. It supplements ISO 243
11、1:1993, Paints and varnishes Determination of flow time by use of flow cups. It specifies the general procedure to be followed in determining the dynamic viscosity of paints, varnishes and related products at a rate of shear between 9 000 s1 and 12 000 s1. The value obtained gives information about
12、the resistance offered by the material to brushing, spraying and roller coating during application. The method specified in this part of ISO 2884 is suitable for all paints and varnishes whether they are Newtonian in behaviour or not. Materials containing dispersions of large particles will produce
13、spurious results. 2 Normative references The following normative documents contain provisions which, through reference in this text, constitute provisions of this part of ISO 2884. For dated references, subsequent amendments to, or revisions of, any of these publications, do not apply. However, part
14、ies to agreements based on this part of ISO 2884 are encouraged to investigate the possibility of applying the most recent editions of the normative documents indicated below. Members of ISO and IEC maintain registers of currently valid International Standards. ISO 1513:1992, Paints and varnishes Ex
15、amination and preparation of samples for testing. ISO 15528:1), Paints and varnishes Sampling. 3 Apparatus 3.1 Viscometer, with cone-and-plate geometry, working at a rate of shear between 9 000 s1 and 12 000 s1. The actual apparatus used shall be agreed between the interested parties, and the detail
16、s given in the test report. A simple apparatus for routine use is described in annex A. 1) To be published. (Revision of ISO 842:1984 and ISO 1512:1991) -,-,- ISO 2884-1:1999(E) ISO 2 4 Sampling Take a representative sample of the product to be tested, as described in ISO 15528. Then examine the sam
17、ple and prepare it for testing, as described in ISO 1513. If the sample has a tendency to sediment or separate on standing, stir it until it is homogeneous, taking care not to incorporate air bubbles. The sample shall be free of any foreign matter or lumps. The volume of the sample shall be sufficie
18、nt to fill the gap between the cone and plate. NOTE Samples containing large particles will produce anomalous results and can cause damage to the instrument. For cone-and-plate viscometers with truncated cones, the larger particles in the sample should be less than a tenth the size of the gap betwee
19、n the cone and the plate. 5 Checking the apparatus Check the apparatus on a regular basis in accordance with the manufacturers recommendations and from experience gained in the use of the instrument, comparing the results against those from previous checks and hence establishing the frequency of che
20、cking required. Check by carrying out the determination as outlined in clause 7, using standard refined mineral oils having Newtonian characteristics and known viscosities (use three mineral oils with viscosities certified by an approved laboratory and lying between 0,05 Pas and 0,5 Pas). Check the
21、cones regularly for wear and replace them if they show signs of scoring or, for cones that are not truncated, signs of flattening. If the readings obtained differ from the known viscosities of the standard oils by more than 5 %, the apparatus shall be checked by a competent instrument engineer or re
22、turned to the manufacturer for adjustment. NOTE Silicone oils should preferably be avoided because of their tendency to contaminate instruments and because of the possibility of shear-thinning behaviour at high shear rates. 6 Checking the temperature control To check that the temperature does not ch
23、ange during the determination, carry out the determination as outlined in clause 7 with the standard oil of the highest viscosity as indicated by the scale reading. Allow the viscometer to run with this oil for 5 min, after which the reading shall not have decreased by more than 10 %. If the decreas
24、e is more than 10 %, the apparatus is unsuitable, as set up, for the determination of viscosities at high rates of shear in accordance with this part of ISO 2884. 7 Procedure 7.1 Carry out the following sequence of operations in duplicate immediately after the preparation of the sample in accordance
25、 with clause 4, first allowing the apparatus to warm up in accordance with the manufacturers instructions. When the viscosities of products are being compared, the rate of shear shall be the same. The determination shall be carried out at (23 0,2) C unless otherwise agreed. 7.2 Adjust the temperatur
26、e of the stationary part of the viscometer to (23 0,2) C or to an alternatively agreed temperature. Transfer a suitable amount of the product to be tested to the appropriate part of the viscometer, taking care to avoid the inclusion of air bubbles, and adjust the other part to the correct position.
27、Wait for the prescribed time, dependent on the apparatus used, to allow the sample to attain the agreed temperature. 7.3 Start the rotor and record the reading on the scale when the reading becomes steady. If the reading does not become steady after 15 s, the reading at 15 s shall be recorded and th
28、e lack of a constant reading shall be mentioned in the test report. 7.4 If the reading does not directly indicate the viscosity, multiply the reading by the appropriate conversion factor or use an appropriate calibration curve to obtain the viscosity. -,-,- ISO ISO 2884-1:1999(E) 3 8 Cleaning the ap
29、paratus Clean the stator and rotor carefully after each determination, employing a suitable solvent. The procedure to be used will depend on the apparatus, but care shall be taken to remove all of the test material and cleaning solvent. Cleaning utensils which may damage the apparatus shall not be u
30、sed. Metal cleaning tools shall never be used. 9 Precision The results of two determinations with the same apparatus taken shortly after one other in the same laboratory by the same operator shall not differ by more than 5 % of their mean. 10 Test report The test report shall contain at least the fo
31、llowing information: a) all details necessary to identify the product tested; b) a reference to this part of ISO 2884 (ISO 2884-1); c) the type of apparatus used, stating the angle and diameter of the cone; d) the shear rate used; e) the temperature used; f) the test results in Pas or mPas; g) any d
32、eviation from the test procedure described, and whether agreed or not; h) the date of the test. -,-,- ISO 2884-1:1999(E) ISO 4 Annex A (informative) Cone-and-plate viscometers A.1 Description of a simple cone-and-plate viscometer for routine use Cone-and-plate viscometers comprise a suitable electri
33、c motor that drives, at a constant rotational speed, a cone whose vertex touches a rigid temperature-controlled plate. The torque is measured either electronically or mechanically. Cone-and-plate viscometers are widely used for routine high-shear viscosity measurements. The geometry is shown in Figu
34、re A.1. The instrument is designed so that the cone and motor assembly can be easily raised, first while the test liquid is being placed on the plate and later to permit thorough cleaning after each measurement. In use, the liquid just fills the narrow gap between the plate and the cone. The main ch
35、aracteristics are described in Table A.1. Dimensions in millimetres Figure A.1 Cone and plate geometry Table A.1 Viscometer characteristics Characteristic Viscosity range 0 to 1 Pas (0 to 10 P) Rotational speed(750 10) rpm Cone diameterDependent on speed, angle and torque selected (typically 24 mm)
36、Cone angle0,5 2 produces a shear rate of 9 000 s1 Shear rate (calculated)9 000 s1 -,-,- ISO ISO 2884-1:1999(E) 5 A.2 Viscometric formulae The equations governing the shear stress and shear rate for a cone-and-plate instrument are given below: Shear rate (s1) = w qtan Shear stress (Pa) = 3 2 3 T rp V
37、iscosity (Pas) = 3 2 3 T r tanq wp where Tis the torque, in newton metres; is the angular velocity, in radians per second; ris the radius of the cone, in metres; is the cone angle, in radians. -,-,- ISO 2884-1:1999(E) ISO 6 Bibliography 1 FINK-JENSEN, P.H., and RAASCHOU NIELSEN, H.K.: “Assessment of
38、 application properties of brushing paints“, (report from the Organic Coatings Section, International Union of Pure and Applied Chemistry), J Paint Techn., 43 (1971), No. 561, pp. 60-67. 2 MONK, C.J.H.: “Routine measurement of the viscosity of paint samples“, JOCCA, 49 (1966), pp. 543-550. -,-,- ISO 2884-1:1999(E) ISO ICS 87.040 Price based on 6 pages -,-,-