1、BRITISH STANDARD BS 6859-2: 1987 Operating tables Part 2: Specification for fixed-pedestal general operating tables UDC 615.478.62 Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 This British Standard, having been prepared und

2、er the direction of the Health Care Standards Committee, was published under the authority of the Board of BSI and comes into effect on 30 November 1987 BSI 09-1999 The following BSI references relate to the work on this standard: Committee reference HCC/33 Draft for comment 81/55107 DC ISBN 0 580 1

3、5903 5 Committees responsible for this British Standard The preparation of this British Standard was entrusted by the Health Care Standards Committee (HCC/-) to Technical Committee HCC/33, upon which the following bodies were represented: Association of Anaesthetists of Great Britain and Ireland Ass

4、ociation of Clinical Pathologists British Medical Association British Surgical Trades Association Incorporated Castor Manufacturers (UK) Association Department of Health and Social Security Department of the Environment (Property Services Agency) Malaysian Rubber Producers Research Association Natio

5、nal Association of Theatre Nurses Royal College of Midwives Royal College of Nursing and National Council for Nurses of the United Kingdom Royal College of Surgeons of England Amendments issued since publication Amd. No.Date of issueComments Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:

6、24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 BSI 09-1999i Contents Page Committees responsibleInside front cover Forewordii 1Scope1 2Dimensions and height1 3General requirements1 4Table side bars2 5Dimensions of side bar accessory clamps2 6Radiographic requirements2 7Finishes of e

7、xposed surfaces of table2 8Mattress2 9Marking2 10Information to be supplied by the manufacturer3 Appendix A Recommendations for transporters for use with removable table tops4 Appendix B Recommendations for radio-translucency of the table top4 Figure 1 Typical fixed-pedestal general operating table4

8、 Figure 2 Some table top positions5 Table 1 Tests and requirements for paint and plastics coating finishes3 Publications referred toInside back cover Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 ii BSI 09-1999 Foreword This

9、 Part of BS 6859 has been prepared under the direction of the Health Care Standards Committee at the request of the Department of Health and Social Security. BS 6859 is published in two Parts, Part 1 specifying requirements for mobile general operating tables. Attention is drawn to BS 5724-2.22 whic

10、h specifies requirements for safety of general operating tables. BS 2891 which specifies requirements for operating table covers and pads is currently being revised. When the revision is published, clause 8 of this Part of BS 6859 will be amended to refer to it. Appendix A gives recommendations for

11、design of transporters which, although outside the scope of this Part of BS 6859, may be used with certain tables which have removable tops. A British Standard does not purport to include all the necessary provisions of a contract. Users of British Standards are responsible for their correct applica

12、tion. Compliance with a British Standard does not of itself confer immunity from legal obligations. Summary of pages This document comprises a front cover, an inside front cover, pages i and ii, pages 1 to 6, an inside back cover and a back cover. This standard has been updated (see copyright date)

13、and may have had amendments incorporated. This will be indicated in the amendment table on the inside front cover. Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 BSI 09-19991 1 Scope This Part of BS 6859 specifies requirement

14、s for fixed-pedestal general operating tables with tops of four or more sections. Fixed-pedestal operating tables intended for use with specialist procedures and accessories to tables are outside the scope of this Part of BS 6859, except that dimensions for accessory clamps have been specified to en

15、sure interchangeability of accessories between tables. NOTEThe titles of the publications referred to in this standard are listed on the inside back cover. 2 Dimensions and height NOTEFigure 1 shows schematic representations of a typical fixed-pedestal general operating table. 2.1 Table top The leng

16、th of the table top shall be not less than 1 850 mm, and the width, including the side bars, shall be not more than 620 mm. 2.2 Trunk section(s) The length of the trunk section(s), i.e. excluding head and leg sections, shall be not less than 840 mm and the width of the trunk sections excluding side

17、bars shall be not less than 470 mm. 2.3 Height With the table top horizontal, the height of the table, excluding the mattress, shall be not less than 650 mm. The height of the table shall be adjustable upwards by not less than 200 mm. 3 General requirements NOTE 1Except for movements of the leg sect

18、ion, all movements of the table should be controllable from or near the usual head end of the table. NOTE 2When assessing compliance with 3.2 and 3.3, the table should be lowered by an operator experienced in the use of operating tables. 3.1 Table top see Figure 1(a) and Figure 1(b) If the table top

19、 is removable from the base, release of a locating or locking device shall be necessary to effect removal. 3.2 Trendelenburg position see Figure 2(a) It shall be possible to lower the table top to a position where the angle between the head end and the horizontal is at least 35 but it shall not be p

20、ossible to lower the table top so that the angle exceeds 45. The time taken to lower the table top to an angle of 35 from the horizontal shall not exceed 8 s from engagement of the lowering mechanism. 3.3 Reverse Trendelenburg position see Figure 2(b) It shall be possible to lower the table top to a

21、 position where the angle between the foot end and the horizontal is at least 35 but it shall not be possible to lower the table top so that the angle exceeds 45. The time taken to lower the table top to an angle of 35 from the horizontal shall not exceed 8 s from engagement of the lowering mechanis

22、m. 3.4 Lateral tilt see Figure 2(c) It shall be possible to tilt the table top laterally to either side to a position where the angle between the table top and the horizontal is at least 15 but does not exceed 20. 3.5 Break-back point see Figure 2(d) and Figure 2(e) At the break-back point, the tabl

23、e top shall be capable of flexion to 130 in any position where all sections are above the horizontal. The table top shall be capable of extension to an angle of not less than 220 between the upper surfaces when all sections are below the horizontal. 3.6 Chair or sitting position It shall be possible

24、 to adjust the sections of the table top to form a chair or sitting position. 3.7 Head section The head section shall be capable of adjustment through any angle between 45 1 downwards and 20 upwards. A quick-release mechanism shall be fitted incorporating a self-locking device and capable of overrid

25、ing the adjustment control. 3.8 Single leg section The leg section on tables with a single leg section shall be adjustable downwards through a minimum angle of 100. 3.9 Divided leg section The two leg sections on tables with a divided leg section shall be capable of rotation outwards in a horizontal

26、 plane to subtend an angle of not less than 100 and each shall also be independently adjustable downwards through a minimum angle of 100. 3.10 Detachable sections 3.10.1 In the case of tables with a single leg section, if the table top is not removable and both head and leg sections are detachable,

27、the head and leg sections shall be interchangeable. 3.10.2 Detachable sections shall not be removable without the release of a locating or locking device. Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 2 BSI 09-1999 3.11 Pede

28、stal unit If the pedestal rotates, a mechanism shall be provided to lock it in any position. 4 Table side bars 4.1 Materials Side bars shall be manufactured either from stainless steel of grade 304S15 or 431S29 as specified in BS 970-1 or from a material of equivalent loadbearing capacity, wear resi

29、stance and corrosion resistance. 4.2 Dimensions The cross section of side bars shall be 31.5mm 6.2 0.15 mm. 4.3 Distance of bar from side of table top The gap between side bars and the side of the table top shall be not less than 10 mm. 5 Dimensions of side bar accessory clamps 5.1 Accessory clamp s

30、ockets 5.1.1 Side bar socket. The dimensions of the socket shall be 32.5mm 7.0mm. 5.1.2 Circular accessory socket. The socket shall be a circle of diameter 16.5mm. 5.1.3 Rectangular accessory socket. The dimensions of the socket shall be 32.5mm 7.0mm. 5.1.4 Square accessory socket. The socket shall

31、be a square with edge length 16.5mm. 5.2 Accessory rods 5.2.1 Circular section. The diameter of the cross section of the rod shall be 16.0mm. 5.2.2 Rectangular section. The dimensions of the cross section of the rod shall be 31.5mm 6.2 0.15 mm. 5.2.3 Square section. The edge length of the cross sect

32、ion of the rod shall be 16.0mm. 6 Radiographic requirements NOTEThe design of the operating table should allow the use of a mobile X-ray image intensifier. 6.1 Space shall be provided immediately below the table top to accommodate an X-ray film cassette with dimensions of 383.5 mm 459.5 mm, so that

33、it can be placed under any part of the patient. 6.2 It shall be possible to load an X-ray film cassette from both the head and foot ends of the table. 6.3 A device shall be provided to enable the X-ray film cassette to be placed in position. 7 Finishes of exposed surfaces of table 7.1 Stainless stee

34、l The stainless steel, including welds shall be polished in accordance with finishes 3A, 3B, 4 or 8 of Table 1 of BS 1449-2:1983. 7.2 Coated steel Applied coatings such as stoving paint or plastics coating shall comply with the requirements given in Table 1 when tested by the appropriate method. 7.3

35、 Chromium and nickel plated steel Chromium and nickel plated steel components shall be plated in accordance with service condition number 3 of BS 1224. 7.4 Aluminium alloy Aluminium alloy components, if anodized, shall comply with grade AA5 of BS 1615, or shall be straight grained to a minimum of 36

36、 grit aluminium oxide, satin finished or bright polished. 7.5 Other materials If exposed surfaces of the table are made from other materials, they shall have a corrosion resistance equal to or better than the materials referred to in 7.1 to 7.4. 8 Mattress The table top shall be covered by a removab

37、le, washable, antistatic mattress segmented or hinged to conform to the size and different configurations of the table top. The mattress shall be attached to the table top and shall not impede the adjustments of the table top or the fitting of accessories. 9 Marking The operating table shall be mark

38、ed with the following: a) the name or trademark of the manufacturer; b) the number and date of this British Standard, i.e. BS 6859-2:19871). 0.25+ 0.5 +0.5 0 0+ 0.5 0+ 0.5 0.5+ 0 0+ 0.5 0.5+ 0 0+ 0.5 0.25+ 0.5 0+ 0.5 1) Marking BS 6859-2:1987 on or in relation to a product represents a manufacturers

39、 declaration of conformity, i.e. a claim by or on behalf of the manufacturer that the product meets the requirements of the standard. The accuracy of the claim is therefore solely the responsibility of the person making the claim. Such a declaration is not to be confused with third party certificati

40、on of conformity, which may also be desirable. Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 BSI 09-19993 Table 1 Tests and requirements for paint and plastics coating finishes 10 Information to be supplied by the manufactur

41、er The manufacturer shall provide a statement of the attenuation of the table top. NOTERecommendations for radio-translucency of the table top are given in Appendix B. TestMethodRequirements 1. Bend testBS 3900-E1 using the type 2 apparatus and a 6 mm mandrel and burnished steel No cracks or loss of

42、 adhesion when examined with normal corrected vision 2. Scratch resistanceBS 3900-E2: Loaded with 2 000 gNo bare metal visible 3. Impact resistanceBS 3900-E3: Indent to 3 mmNo loss of adhesion 4. Cupping testBS 3900-E4: Indent to 3 mm on paint and reverse side No loss of adhesion or cracks when exam

43、ined with normal corrected vision Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 4 BSI 09-1999 Appendix A Recommendations for transporters for use with removable table tops A.1 Transfer of the table top between the pedestal a

44、nd transporter should be possible by manual control. A.2 The table top should be locked into the pedestal on takeover of the table top from the transporter. A.3 Any prepositioning of the patient or addition and fixation of accessories should not affect the stability of a transporter/top combination.

45、 A.4 Provision should be made to immobilize the transporter. A.5 To prevent damage to floors and to ensure ease of manoeuvring, the transporter should be fitted with large diameter wheels. A.6 Table tops, whilst located on their transporter, should be capable of achieving a Trendelenburg position se

46、e Figure 2(a) of not less than 15. A.7 The transporter should have cot sides. Appendix B Recommendations for radio-translucency of the table top The table top, excluding the mattress, should have an attenuation equivalent to no greater than 1 mm thickness of aluminium, determined from measurements o

47、f air Kerma under narrow beam conditions. The measurements should be made with X-radiation from an X-ray source assembly operated at an X-ray tube voltage of 100 kV with a total filtration of 2.5 mm aluminium. The percentage ripple of the X-ray tube voltage should be less than 10 %. NOTEThe break-ba

48、ck point need not be at the centre of the table top. (a) Side view (b) Top view Figure 1 Typical fixed-pedestal general operating table Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 BSI 09-19995 Figure 2 Some table top posit

49、ions Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI 6 blank Licensed Copy: sheffieldun sheffieldun, na, Tue Dec 05 01:24:55 GMT+00:00 2006, Uncontrolled Copy, (c) BSI BS 6859-2:1987 BSI 09-1999 Publications referred to BS 970, Specification for wrought steels for mechanical and allied engineering purposes. BS 970-1, General inspection and testing procedures and specific requirements for carbon, carbon manganese, alloy and stainless steels. BS 122