【BS英国标准】BS EN 866-1-1997 Biological systems for testing sterilizers and sterilization processes Part 1. General requirements.doc

上传人:椰子壳 文档编号:3619986 上传时间:2019-09-18 格式:DOC 页数:17 大小:551.50KB
返回 下载 相关 举报
【BS英国标准】BS EN 866-1-1997 Biological systems for testing sterilizers and sterilization processes Part 1. General requirements.doc_第1页
第1页 / 共17页
【BS英国标准】BS EN 866-1-1997 Biological systems for testing sterilizers and sterilization processes Part 1. General requirements.doc_第2页
第2页 / 共17页
【BS英国标准】BS EN 866-1-1997 Biological systems for testing sterilizers and sterilization processes Part 1. General requirements.doc_第3页
第3页 / 共17页
【BS英国标准】BS EN 866-1-1997 Biological systems for testing sterilizers and sterilization processes Part 1. General requirements.doc_第4页
第4页 / 共17页
【BS英国标准】BS EN 866-1-1997 Biological systems for testing sterilizers and sterilization processes Part 1. General requirements.doc_第5页
第5页 / 共17页
亲,该文档总共17页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

《【BS英国标准】BS EN 866-1-1997 Biological systems for testing sterilizers and sterilization processes Part 1. General requirements.doc》由会员分享,可在线阅读,更多相关《【BS英国标准】BS EN 866-1-1997 Biological systems for testing sterilizers and sterilization processes Part 1. General requirements.doc(17页珍藏版)》请在三一文库上搜索。

1、BRITISH STANDARDBiological systems for testing sterilizers and sterilization processesPart 1. General requirementsThe European Standard EN 866-1 : 1997 has the status of aBritish StandardICS 11.080| BS EN866-1 : 1997|NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAWBS EN 866-1 :

2、 1997Committees responsible for thisBritish StandardThe preparation of this British Standard was entrusted by Technical Committee LBI/35, Sterilizers, autoclaves and disinfectors, to Subcommittee LBI/35/3, Sterilization indicators, upon which the following bodies were represented:ABHI Special Intere

3、st Section (Sterilizers and Disinfectors) Association of British Healthcare IndustriesBLWA Ltd. (the Association of the Laboratory Supply Industry) Department of Health (Medical Devices Agency)European Sterilization Packaging Association (UK) Medical Sterile Products AssociationPublic Health Laborat

4、ory ServiceRoyal College of PathologistsRoyal Pharmaceutical Society of Great BritainThis British Standard, having been prepared under the direction of the Sector Board for Materials and Chemicals, was published under the authority of the Standards Board and comes into effect on15 November 1997BSI 1

5、997The following BSI references relate to the work on this standard:Committee reference LBI/35/3Draft for comment 95/121171 DCISBN 0 580 28377 1Amd. No.DateText affectedAmendments issued since publicationBS EN 866-1 : 1997National forewordThis Part of BS EN 866 has been prepared by Subcommittee LBI/

6、35/3 and is the English language version of EN 866-1 : 1997 Biological systems for testing sterilizers and sterilization processes Part 1: General requirements, published by the European Committee for Standardization (CEN).Cross-referencesPublication referred to Corresponding British StandardEN 550

7、: 1994BS EN 550 : 1994 Sterilization of medical devices Validation and routine control of ethylene oxide sterilizationEN 552 : 1994BS EN 552 : 1994 Sterilization of medical devices Validation and routine control of sterilization by irradiationEN 554 : 1994BS EN 554 : 1994 Sterilization of medical de

8、vices Validation and routine control of sterilization by moist heatEN 868-1 : 1997BS EN 868 Packaging materials and systems for medical devices which are to be sterilizedPart 1 : 1997 General requirements and test methodsEN 1174-1 : 1996BS EN 1174 Sterilization of medical devices Estimation of the p

9、opulation of micro-organisms on productPart 1 : 1996 RequirementsEN 28601 : 1992BS EN 28601 : 1992 Specification for representation of dates and times in information interchangeCompliance with a British Standard does not of itself confer immunity from legal obligations.Summary of pagesThis document

10、comprises a front cover, an inside front cover, pages i and ii, the ENtitle page, pages 2 to 10, an inside back cover and a back cover.BSI 1997iiiblankEUROPEAN STANDARDEN 866-1NORME EUROPE ENNEEUROPA ISCHE NORMFebruary 1997ICS 11.080Descriptors: Medical equipment, sterilizers, bioassay, manufacturin

11、g, biological indicators, specifications, preparation, packing, labellingEnglish versionBiological systems for testing sterilizers and sterilization processes Part 1: General requirementsSyste mes biologiques pour lessai des ste rilisateurs et les proce de s de ste rilisation Partie 1: Exigences ge

12、ne ralesBiologische Systeme fu r die Pru fung von Sterilisatoren und Sterilisationsverfahren Teil 1: Allgemeine AnforderungenThis European Standard was approved by CEN on 1997-01-10. CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving t

13、his European Standard the status of a national standard without any alteration.Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the Central Secretariat or to any CEN member.This European Standard exists in three official versions (E

14、nglish, French, German). A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the Central Secretariat has the same status as the official versions.CEN members are the national standards bodies of Austria, Belgium, Denmark,

15、 Finland, France, Germany, Greece, Iceland, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and United Kingdom.CENEuropean Committee for Standardization Comite Europe en de Normalisation Europa isches Komitee fu r NormungCentral Secretariat: rue de Stassart 36,

16、B-1050 Brussels1997 Copyright reserved to CEN membersRef. No. EN 866-1 : 1997 EPage 9EN 866-1 : 1997BSI 1997ForewordThis European Standard has been prepared by Technical Committee CEN/TC 102, Sterilizers for medical purposes, the secretariat of which is held by DIN.This Standard is one of a series o

17、f European Standards concerned with biological systems for testingsterilizers. These European Standards are:EN 866-1Biological systems for testing sterilizers and sterilization processes Part 1: General requirementsEN 866-2Biological systems for testing sterilizers and sterilization processes Part 2

18、: Particular systems for use in ethylene oxide sterilizersEN 866-3Biological systems for testing sterilizers and sterilization processes Part 3: Particular systems for use in moist heat sterilizersprEN 866-4Biological systems for testing sterilizers and sterilization processes Part 4: Particular sys

19、tems for irradiation sterilizersprEN 866-5Biological systems for testing sterilizers and sterilization processes Part 5: Particular systems for low- temperature steam formaldehyde sterilizersprEN 866-6Biological systems for testing sterilizers and sterilization processes Part 6: Particular systems f

20、or dry heat sterilizersprEN 866-7Biological systems for testing sterilizers and sterilization processes Part 7: Particular requirements for self-contained biological indicator systems for use in moist heat sterilizersprEN 866-8Biological systems for testing sterilizers and sterilization processes Pa

21、rt 8: Particular requirements for self-contained biological indicator systems for use in ethylene oxide sterilizersIn addition CEN/TC 102 Working Group 7 has prepared a series of European Standards describingnon-biological indicators for use in sterilizers. TheseEuropean Standards are:EN 867-1Non-bi

22、ological systems for use in sterilizers Part 1: General requirementsEN 867-2Non-biological systems for use in sterilizers Part 2: Process indicators (Class A)EN 867-3Non-biological systems for use in sterilizers Part 3: Specification for Class B indicators for use in the Bowie and Dick testThis Euro

23、pean Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by August 1997, and conflicting national standards shall be withdrawn at the latest by August 1997.This European Standard has been prepared under a mandate give

24、n to CEN by the European Commission and the European Free Trade Association, and supports essential requirements of EU Directive(s).According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austri

25、a, Belgium, Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and the United Kingdom.ContentsPageForeword 2Introduction 31 Scope 32 Normative references 33 Definitions 34 General requirements 45 Test organisms an

26、d their preparation for the manufacture of inoculatedcarriers 56 Carriers and their inoculation 57 Packaging and labelling of inoculatedcarriers for distribution 68 Preparation of biological indicatorsfrom inoculated carriers 69 Packaging and labelling of biologicalindicators 610 Determination of te

27、st organismresistance 7AnnexesA (normative) Determination of growth inhibition by carriers exposed to thesterilization process 8B (normative) Determination of testorganism resistance 8C (informative) Relationship between the components of biological testsystems 10D (informative) Bibliography 10Intro

28、ductionEuropean Standards for sterilizers (EN 285 andprEN 1422) and for the validation and process control of sterilization (EN 550 and EN 554) describe performance tests for sterilizers and methods of validation and routine control, respectively.This standard specifies the general requirements for

29、biological indicators, and subsequent Parts specify the particular requirements for biological indicators for defined sterilization processes. The use of the indicators specified in this standard is described inEN 550 and EN 554.The use of a biological system for testing a sterilization process does

30、 not allow necessarily the same level of sensitivity in response to inadequate levels of all the critical variables of the process.The biological indicators specified in this standard are not intended for use in any process other than that stated by the manufacturer. The use of a biological indicato

31、r in a process other than that stated by the manufacturer can give dangerously misleading results.The performance of a biological indicator can be affected by the conditions of storage prior to use, the methods of use, and the techniques employed after exposure to the process. For these reasons, the

32、 recommendations of the manufacturer for storage and use should be followed and biological indicators should be transferred to the specified recovery conditions as soon as possible after exposure to the process. Biological indicators should not be used beyond any expiry date stated by the manufactur

33、er.Biological indicators should always be used in combination with appropriate physical and/or chemical monitoring in demonstrating the efficacy of a sterilization process. When a physico-chemical variable of a sterilization process is outside its specified limits,a sterilization cycle should always

34、 be regarded as unsatisfactory, irrespective of the results obtained from the biological indicators. See also EN 550 and EN 554.1 ScopeThis Part of EN 866 specifies general requirements for the manufacture of biological systems to be used in testing sterilizers and sterilization processes.The requir

35、ements of this Part of EN 866 apply to all biological systems specified in subsequent Parts ofEN 866, unless the requirement is modified or added to by a subsequent Part, in which case the requirement of the particular Part will apply.2 Normative referencesThis European Standard incorporates, by dat

36、ed or undated reference, provisions from other publications. These normative references are cited at the appropriate places in the text and the publications are listed hereafter. For dated references, subsequentamendments to or revisions of any of these publications apply to this European Standard o

37、nlywhen incorporated in it by amendment or revision. For undated references, the latest edition of the publication referred to applies.EN 550Sterilization of medical devices Validation and routine control of ethylene oxide sterilizationEN 552Sterilization of medical devices Validation and routine co

38、ntrol of sterilization by irradiationEN 554Sterilization of medical devices Validation and routine control of sterilization by moist heatEN 868-1Packaging materials and systems for medical devices which are to be sterilized Part 1: General requirements Requirements and test methodsEN 1174-1Steriliza

39、tion of medical devices Estimation of the population of micro-organisms on product Part 1: RequirementsEN 28601 : 1992 Data elements and interchange formats Information interchange Representation of dates and times(ISO 8601 : 1988 and technical corrigendum 1 : 1991)3 DefinitionsFor the purposes of t

40、his standard, the following definitions apply.3.1 biological indicatorAn inoculated carrier contained within its primary pack ready for use (see figure C.1 in annex C).3.2 biological systemsThose systems which depend for their function on the demonstration of viability of a test organism.NOTE. This

41、demonstration of viability can be achieved by any method suitable for indicating, qualitatively or quantitatively, either the metabolic activity or the replication of the organism. Tests based on physical or chemical detection of change in a particular chemical entity, whether or not this was origin

42、ally derived or isolated from a biological system (e.g. an enzyme) are consideredin the series EN 867 Non-biological systems for use in sterilizers.3.3 carrierThe supporting material on which the test organisms are deposited (see figure C.1 in annex C).3.4 culture collection numberThe unique identif

43、ication allocated by the recognized culture collection.3.5 culture conditionsThe manufacturers stated combination of conditions, including the growth medium with the period and temperature of incubation, used to promote germination, outgrowth and/or multiplication of the test organism.3.6 D value (d

44、ecimal reduction value)The time in minutes, or the absorbed irradiation dose in kilograys, required to secure inactivation of 90 % of the test organisms under stated exposure conditions.3.7 inactivationThe loss of the ability of the test organisms to germinate, outgrow and/or multiply under culture conditions.3.8 inoculated carrierA carrier on which a defined number